Coefficient of variation calculator
Sample standard deviation, divided by n−1. CV% is that SD divided by the mean. A full precision study still needs a design such as CLSI EP05.
How to use the coefficient of variation calculator
Paste replicate results separated by commas, spaces or semicolons, then read mean, sample SD and CV%. Estimate detection limits with the LoD estimator, trend QC rules with the Westgard QC helper, or summarize clinical 2×2 performance with the sensitivity and specificity calculator.
Planning assay precision or IVD analytical studies?
A single CV% is not a precision claim. Book a Consultation with our clinical / regulatory team to design repeatability and reproducibility studies (for example CLSI EP05) and tie them to your device’s intended use.
Book a ConsultationFAQ
- What is the coefficient of variation?
- The coefficient of variation (CV) is the standard deviation divided by the mean, usually expressed as a percent. It lets you compare relative precision across measurements that have different units or concentration levels.
- Is this the sample or the population standard deviation?
- Sample. The standard deviation divides by n−1 (Bessel’s correction). That is the usual choice for a finite set of assay replicates rather than an entire population.
- Does a CV% replace a precision study such as CLSI EP05?
- No. Repeatability and within-laboratory or multi-site reproducibility still need a designed experiment. CLSI EP05 is a commonly cited evaluation-of-precision protocol for quantitative measurement procedures. This page is the first arithmetic check on a paste of replicates.
- How does CV relate to LoD and clinical claims?
- Precision often worsens near the limit of detection or quantitation. A CV that looks acceptable at a mid-range control level may not support claims near LoD. Pair precision work with LoD studies (for example CLSI EP17) when those claims matter to the intended use.
- Should I pool replicates across days or instruments?
- Only when the protocol says so. Mixing days, lots, operators or instruments without a design confounds repeatability with intermediate precision or reproducibility. Keep the paste homogeneous with the precision component you are estimating.
- What if the mean is near zero?
- CV is undefined or unstable when the mean approaches zero, because you divide by the mean. Use another precision metric or a different concentration level for those data.
Educational only. Not medical advice, not a regulatory determination, and not a substitute for qualified RA/QA, legal, clinical, or statistical counsel. Coefficient of variation does not submit to FDA, Notified Bodies, or IRBs.